What DCD does
Before (and on Pro Plus, around) each pulse, the DCD sprays a measured burst of cryogen that cools the skin's surface. The system meters that burst precisely; anything that disturbs it — canister, valve, nozzle, timing — shows up as a fault or a complaint.
Common DCD complaints
- Cooling bursts feel weak or uneven between pulses
- Faults pointing at the cryogen canister or delivery
- Spray audible but treatment area not cooling
- Issues that start right after a canister change
- Frosting patterns or residue at the handpiece window
Likely causes
A canister not seated or not the correct type, a valve or nozzle partly blocked, a delivery line issue, or a calibration problem in how the burst is timed against the pulse. Window condition matters too: a fouled window changes both cooling and energy delivery.
What we do
We verify canister and seating, test burst consistency, inspect the nozzle and window, and measure delivered energy with cooling active — because DCD faults and output faults often travel together on this platform.
FAQ
Questions on this topic
The problem started when we changed the canister. Related?
Very often, yes. Seating and canister type are the first checks — tell your technician the timing; it is a real diagnostic clue.
Can we treat without DCD while waiting for service?
No. The cooling is part of the treatment's safety design on this platform. A DCD fault is a stop-and-call fault.