What calibration actually is
A technician measures the energy the system truly delivers — at the handpiece, across the settings and spot sizes in use — and adjusts the system until screen and reality agree within the manufacturer's tolerance. It is measurement plus adjustment, not a button in a menu.
When a system needs it
- After a lamp, diode, tube or board replacement
- After fiber, handpiece or arm work
- When output checks drift outside tolerance
- When treatment results change without a settings change
- On the manufacturer's schedule, as part of preventive maintenance
What you receive
A record of measured values before and after, per wavelength and handpiece where relevant. That record is part of the machine's service history — keep it with the device file.
Calibration versus repair
Calibration corrects drift. It cannot resurrect a worn lamp or a dying diode — a system that cannot reach tolerance needs the underlying repair first, and an honest technician will tell you which one you are buying.
FAQ
Questions on this topic
Can I verify calibration myself?
Built-in energy checks are a useful smoke test, but verification needs a calibrated external meter at the handpiece. That is the measurement that counts.
How long does calibration take?
Typically part of a single on-site visit for one platform, longer when several handpieces or wavelengths are involved.