Safety first, always
A laser that will not fire is often protecting you or itself. Interlocks — handpiece seated, fiber connected, footswitch circuit complete, room door switch where fitted — all sit in series: any one open and the system stays silent. That silence is information, not attitude.
The usual causes
- A handpiece or fiber not fully seated or not recognized
- A footswitch or its cable damaged — the most stepped-on part in the room
- A safety interlock held open by a cover, connector or sensor
- A fault the system logged earlier that now blocks firing
- Calibration or self-test incomplete after a lamp or board event
What you can check safely
Reseat the handpiece and fiber until they click. Swap the footswitch if a spare exists. Note any code on screen and whether firing fails on every handpiece or only one — that single detail halves the diagnosis. Do not open covers and do not defeat interlocks: a firing fault you bypass can become an injury.
When to call
If the system will not fire after reseating and a restart, or if any code accompanies the silence, call with the model, the code, and which handpieces are affected.
FAQ
Questions on this topic
The laser fires on one handpiece but not another. What does that mean?
The fault follows the handpiece — its connector, recognition contacts or internal switch — not the platform. Tell your technician; it shortens the visit.
Can a dead footswitch really stop everything?
Yes. It is a series safety device by design, and its cable lives on the floor. It is the first thing we test and the cheapest fix on this list.